MEAS
MEM
LOG
SET
Welcome
1.The meter is capable of adjusting temperature value to
get more accurate reading .
oo
2.The adjustable temperature range is from -5 C to 80 C,
enter a value out of the range, "OVER RANGE" will show on
the screen under Temp. Press ENTER under over range
o
,TEMP will show a default "-5 C".
3."+/-" key is for changing positive value to negative value.
Volt:
EXIT
TEMP
CAL
PRN
Temp:
PH:
28C
7.60
-34.9mV
NonATC
Volt:
EXIT
TEMP
CAL
PRN
Temp:
PH:
28C
7.60
-34.9mV
NonATC
Temp: -5.0
o
C
EXIT
EXIT
LXW
LXW
ENTER
ENTER
+ / -
+ / -
CLR
CLR
EXIT
LXW
ENTER
+ / -
CLR
Note 1:
This is to calibrate under NonATC
( No Temperature probe is plugged
side by pH probe )
Note 2:
If you always see "--.--" pH in auto-
juddge, please refer to trouble
shooting.
Note 3: Display turns to normal
mode if non-operation during
Temp. is flashing for editing.
9861/9661 pH/mV Probe Calibration procedure-3
Temp: 18.2
o
C
RANGE:-5.0~80.0C
Temp: -18.2
o
C
RANGE:-5.0~80.0C
RANGE:-5.0~80.0C
OVER RANGE
27