9135Infrared Thermometer CalibratorUser’ s GuideRev. 5A2604Hart Scientific

Pagina 2 - Rev. 5A2604

2 IntroductionHart Scientific’s Model 9135 Infrared Thermometer Calibrator is a small, por-table instrument designed for quick on-site checks and cali

Pagina 3 - Table of Contents

3 Specifications and EnvironmentalConditions3.1 SpecificationsSet-points 50°C (122°F), 100°C (212°F), and 150°C (302°F)Accuracy ±1.0°CStability (2-sig

Pagina 4

3.3 WarrantyFluke Corporation, Hart Scientific Division (Hart) warrants this product to befree from defects in material and workmanship under normal u

Pagina 5 - 1 Before You Start

4 Quick Start4.1 UnpackingUnpack the calibrator carefully and inspect it for any damage that may have oc-curred during shipment. If there is shipping

Pagina 6 - 1.2 Safety Information

The mains supply must be reasonably stable and free from surges, dropouts,and interference that can affect the temperature stability of the IR target.

Pagina 7 - Cautions

5 Parts and ControlsThe user should become familiar with the 9135 calibrator and its parts. Refer toFigures 1 and 2 as you read the following sections

Pagina 8

5.4 Right Side PanelThe right side panel is illustrated in Figure 2.Test Well - The right side panel has a 6.3 mm (0.25") hole near its center th

Pagina 9 - Authorized Service Centers

adjust these potentiometers unless qualified to do so as part of a regular mainte-nance and calibration schedule (see Section 9).155 Parts and Control

Pagina 10 - 2 Introduction

6 General OperationThis section explains basic operation of the 9135 calibrator.6.1 Set-upPlace the calibrator on a flat surface with at least three i

Pagina 11 - Conditions

7 Controller OperationThe 9135 calibrator uses a precise analog proportional-integral controller thatmaintains the temperature of the IR target. The s

Pagina 12 - 3.3 Warranty

Rev. 5A2604Fluke Corporation, Hart Scientific Division799 E. Utah Valley Drive • American Fork, UT 84003-9775 • USAPhone: +1.801.763.1600 • Telefax: +

Pagina 13 - 4 Quick Start

8 Thermometer CalibrationThe 9135 calibrator was designed primarily for calibrating infrared thermome-ters. However, it can also be used to calibrate

Pagina 14 - User’s Guide

bration. DO NOT attempt to calibrate thermometers unless the greenready LED is lit, otherwise accuracy may be compromised.•The 9135’s test well is int

Pagina 15 - 5 Parts and Controls

9 Calibration ProcedureThe 9135 should be calibrated regularly to ensure accuracy. A one-year or lesscalibration interval is recommended. Calibration

Pagina 16 - 5.4 Right Side Panel

10 Maintenance•This calibration instrument has been designed for a minimum of requiredmaintenance as long as it is handled with proper care. Please re

Pagina 17 - Calibration

WARNING: Never introduce any fluids or other foreign material onto theIR target or into the well block. This will damage the calibrator, cause lossof

Pagina 18 - 6 General Operation

11 TroubleshootingIf problems arise while operating the 9135, this section provides some sugges-tions that may help you solve the problem. If, after c

Pagina 19 - 7 Controller Operation

Table of Contents1 Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1 Symbols Used . . . . . . .

Pagina 20 - 8 Thermometer Calibration

9 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . 2310 Maintenance . . . . . . . . . . . . . . .

Pagina 21

1 Before You Start1.1 Symbols UsedTable 1 lists the International Electrical Symbols. Some or all of these symbolsmay be used on the instrument or in

Pagina 22 - 9 Calibration Procedure

Symbol DescriptionCanadian Standards AssociationOVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 re-fers to the level of Impul

Pagina 23 - 10 Maintenance

•The instrument can generate high temperatures. Precautions must betaken to prevent personal injury or damage to objects. Probes may be veryhot when r

Pagina 24

American Fork, UT 84003-9775USAPhone: +1.801.763.1600Telefax: +1.801.763.1010E-mail: [email protected] Nederland B.V.Customer Support Se

Pagina 25 - 11 Troubleshooting

Telefax: +65 6799-5588E-mail: [email protected] contacting these Service Centers for support, please have the followinginformation available:•